Floor / Pallet Scale - 9 -
Press [FUN] to confirm and go to label format selection. Options: 1/2
P11 FL: Weight adjustment for acceleration of gravity change.
P12 KB: on: keyboard beep is on; off: keyboard beep is off
P13 UT: Unit Setting.
Default unit: lb/kg
Lb on/off
U1 on/off U2 on/off (Unit 1 or Unit 2 setting)
P14 Pn: Password Setting.
P1: Enter new password Press [FUN] to confirm.
P2: Re-enter the new password Press [FUN] to confirm. Screen will show done.
Chapter 5 Calibration
Enter Parameter setting P8.
Unload: Empty the scale platform. Wait until the STAB sign is on, press [FUN].
Input the calibration weight value and press [FUN]
Load: Place the calibration weight on the platform, wait until the STAB sign is on, press [FUN]
The indicator will go to weighing mode when the calibration is completed.
Chapter 6 Error Indication
--OL--
The scale is overloaded. Empty the scale platform.
Err 1 / Err 2
Date and Time are not right. Reset the date and time.
Err 4
The zero point is out of range upon powering-on.